Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Amprobe AM-1280: (1 year) CAL VER /5520 DATE: 25-Aug-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 54 NUMBER OF LINES: 180 CONFIGURATION: Fluke 5520A ============================================================================= # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F V 1.002 ASK+ R K X 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Warm-up time: 30 min. - REQUIRED 1.004 DISP Temperature: 20 - 30 degC 1.004 DISP Humidity: <80% 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 HEAD {DC VOLTAGE} 1.007 DISP Connect the 5520A to the UUT as follows: 1.007 DISP [32] 5520A NORMAL HI to VOLT/OHM/ADP 1.007 DISP [32] 5520A NORMAL LO to COM 1.008 HEAD DC VOLTAGE: {400mV Range} 1.009 5520 400 300.0mV 0.3% 0.2U 2W 2.001 HEAD DC VOLTAGE: {4V Range} 2.002 5520 4 3.000V 0.3% 0.002U 2W 3.001 HEAD DC VOLTAGE: {40V Range} 3.002 5520 40 30.00V 0.3% 0.02U 2W 4.001 HEAD DC VOLTAGE: {400V Range} 4.002 5520 400 300.0V 0.3% 0.2U 2W 5.001 HEAD DC VOLTAGE: {1000V Range} 5.002 5520 1000 900V 0.3% 2U 2W 6.001 HEAD {AC VOLTAGE} 6.002 HEAD AC VOLTAGE: {400mV Range} 6.003 5520 400 300.0mV 1% 0.5U 60H SI 2W 7.001 5520 400 300.0mV 1.5% 0.5U 100H SI 2W 8.001 5520 400 300.0mV 1.5% 0.5U 1kH SI 2W 9.001 HEAD AC VOLTAGE: {4V Range} 9.002 5520 4 3.000V 1% 0.005U 60H SI 2W 10.001 5520 4 3.000V 1% 0.005U 100H SI 2W 11.001 5520 4 3.000V 1% 0.005U 1kH SI 2W 12.001 HEAD AC VOLTAGE: {40V Range} 12.002 5520 40 30.00V 1% 0.05U 60H SI 2W 13.001 5520 40 30.00V 1% 0.05U 100H SI 2W 14.001 5520 40 30.00V 1% 0.05U 1kH SI 2W 15.001 HEAD AC VOLTAGE: {400V Range} 15.002 5520 400 300.0V 1% 0.5U 60H SI 2W 16.001 5520 400 300.0V 1% 0.5U 100H SI 2W 17.001 5520 400 300.0V 1% 0.5U 1kH SI 2W 18.001 HEAD AC VOLTAGE: {750V Range} 18.002 5520 750 700V 1% 5U 60H SI 2W 19.001 5520 750 700V 1% 5U 100H SI 2W 20.001 5520 750 700V 1% 5U 1kH SI 2W 21.001 HEAD {RESISTANCE} 21.002 DISP Connect the 5520 to the UUT as follows: 21.002 DISP [32] 5520A AUX HI to UUT V/OHM 21.002 DISP [32] 5520A AUX LO to UUT COM 21.002 DISP [32] 5520A NORMAL HI to UUT V/OHM 21.002 DISP [32] 5520A NORMAL LO to UUT COM 21.002 DISP Connect the SENSE leads DIRECTLY to the UUT. Connect 21.002 DISP the NORMAL leads to the SENSE leads AT the UUT. 21.003 HEAD RESISTANCE: {400 Ohm Range} 21.004 5520 400 390.0Z 0.7% 0.4U CW 22.001 HEAD RESISTANCE: {4 kOhm Range} 22.002 5520 4 3.900kZ 0.7% 0.003U CW 23.001 HEAD RESISTANCE: {40 kOhm Range} 23.002 5520 40 39.00kZ 0.7% 0.04U CW 24.001 HEAD RESISTANCE: {400 kOhm Range} 24.002 DISP Connect the 5520 to the UUT as follows: 24.002 DISP [32] 5520A NORMAL HI to UUT V/OHM 24.002 DISP [32] 5520A NORMAL LO to UUT COM 24.003 5520 400 390.0kZ 0.7% 0.4U 2W 25.001 HEAD RESISTANCE: {4 MOhm Range} 25.002 5520 4000 3900kZ 0.7% 4U 2W 26.001 HEAD RESISTANCE: {40 MOhm Range} 26.002 5520 40 39.00MZ 1.5% 0.05U 2W 27.001 HEAD {CONTINUITY} 27.002 DISP Select the Continuity function 27.003 HEAD CONTINUITY: {400 Ohms} 27.004 5520 400.0Z S CW 27.005 EVAL Is {continuity alarm off}? 28.001 HEAD CONTINUITY: {<40 Ohms} 28.002 5520 35.0Z S CW 28.003 EVAL Is {continuity alarm sounding}? 29.001 HEAD {DIODE} 29.002 DISP Select the Diode funcition 29.003 5520 0V S 2W 29.004 EVAL [N]Does {display indicate an overload condition}? 30.001 5520 0.700V 2% 2W 31.001 HEAD {CAPACITANCE} 31.002 HEAD CAPACITANCE: {4nF Range} 31.003 5520 4 3.000nF 1% 0.005U 2W 32.001 HEAD CAPACITANCE: {40nF Range} 32.002 5520 40 30.00nF 1% 0.05U 2W 33.001 DISP Connect the 5520 to the UUT as follows: 33.001 DISP [32] 5520A AUX HI to UUT V/OHM 33.001 DISP [32] 5520A AUX LO to UUT COM 33.001 DISP [32] 5520A NORMAL HI to UUT V/OHM 33.001 DISP [32] 5520A NORMAL LO to UUT COM 33.001 DISP Connect the SENSE leads DIRECTLY to the UUT. Connect 33.001 DISP the NORMAL leads to the SENSE leads AT the UUT. 33.002 HEAD CAPACITANCE: {400 nF Range} 33.003 5520 400 300.0nF 1% 0.5U CW 34.001 HEAD CAPACITANCE: {4 uF Range} 34.002 5520 4 3.000uF 1% 0.005U CW 35.001 HEAD CAPACITANCE: {40 uF Range} 35.002 5520 40 30.00uF 3% 0.05U CW 36.001 HEAD {ADP} 36.002 DISP Connect the 5520 to the UUT as follows: 36.002 DISP [32] 5520A NORMAL HI to UUT V/OHM/ADP 36.002 DISP [32] 5520A NORMAL LO to UUT COM 36.003 DISP Select the ADP function. 36.004 5520 10mV S 2W 36.005 EVAL Does UUT {indicate 96 to 104 counts}? 37.001 ASK- R 37.002 HEAD {FREQUENCY} 37.003 DISP Select the Frequency function. 37.004 HEAD FREQUENCY: {100 Hz Range} 37.005 5520 100 90.00H 0.2% 0.04U 1Vpp SI 2W 38.001 HEAD FREQUENCY: {1 kHz Range} 38.002 5520 1000 900.0H 0.2% 0.4U 1Vpp SI 2W 39.001 HEAD FREQUENCY: {10 kHz Range} 39.002 5520 10 9.000kH 0.2% 0.004U 1Vpp SI 2W 40.001 HEAD FREQUENCY: {100 kHz Range} 40.002 5520 100 90.00kH 0.2% 0.04U 1Vpp SI 2W 41.001 HEAD FREQUENCY: {1 MHz Range} 41.002 5520 1000 900.0kH 0.2% 0.4U 1Vpp SI 2W 42.001 ASK+ R 42.002 HEAD {DC CURRENT} 42.003 RNG 4mA 42.004 DISP Connect the 5520 to the UUT as follows: 42.004 DISP [32] 5520A AUX HI to UUT mA 42.004 DISP [32] 5520A AUX LO to UUT COM 42.005 HEAD DC CURRENT: {4 mA Range} 42.006 5520 4 3.000mA 0.5% 0.002U 2W 43.001 HEAD DC CURRENT: {40 mA Range} 43.002 5520 40 30.00mA 0.5% 0.02U 2W 44.001 HEAD DC CURRENT: {400 mA Range} 44.002 5520 400 300.0mA 0.5% 0.2U 2W 45.001 HEAD {AC CURRENT} 45.002 HEAD AC CURRENT: {4 mA Range} 45.003 5520 4 3.000mA 1.2% 0.005U 40H SI 2W 46.001 5520 4 3.000mA 1.2% 0.005U 1kH SI 2W 47.001 HEAD AC CURRENT: {40 mA Range} 47.002 5520 40 30.00mA 1.2% 0.05U 40H SI 2W 48.001 5520 40 30.00mA 1.2% 0.05U 1kH SI 2W 49.001 HEAD AC CURRENT: {400 mA Range} 49.002 5520 400 300.0mA 1.2% 0.5U 40H SI 2W 50.001 5520 400 300.0mA 1.2% 0.5U 1kH SI 2W 51.001 HEAD AC CURRENT: {10 A Range} 51.002 RNG 10A AC 51.003 DISP Connect the 5520 to the UUT as follows: 51.003 DISP [32] 5520A 20A to UUT 10A 51.003 DISP [32] 5520A AUX LO to UUT COM 51.004 5520 10 10.00A 1.5% 0.05U 60H SI HC 2W 52.001 5520 10 10.00A 1.2% 0.05U 1kH SI HC 2W 53.001 HEAD {DC CURRENT} 53.002 5520 10 10.00A 1% 0.03U HC 2W 54.001 END